Instrument
Altec, Beckman, Biorad, Dupont, Erma
Hitachi, Jasco, Unicam, Varian
Electrical
Filament, Heater Voltage |
Start |
10 VDC |
Operating |
5 - 7 VDC |
Filament, Heater Current |
1.5A Max |
Discharge, Arc Voltage |
Strike |
@25°C >350 VDC |
Operating |
60 - 90 VDC |
Discharge, Arc Current |
10 -300mA |
Leads |
Black |
Filament & Cathode |
Red |
Anode |
Terminals |
Spade Terminal |
Optical
Window material |
UV Transmitting Borosilicate |
Source aperture |
Ø 1mm Diameter |
Wavelength range |
185nm - 400nm |
Noise |
<1 x 10-4 Absorption units |
Drift |
< 3 x 10-3 Absorption units |
Operational Life
When radiant energy reaches 50% of the initial energy at 300mA(dc).
Operational Life
When radiant energy reaches 50% of the initial energy at 300mA(dc).
Warranty:
- Do not subject the lamp to mechanical shock.
- Before using the lamp, ensure glass envelope is clean and free from dirt. Envelope may be gently wiped with either alcohol or acetone.
Caution: the circumference of the glass envelope around the light box is very thin and should be handled with extreme care.
- Care must be taken not to touch lamp body with bare hand, as it's temperature reaches over 200°C during operation.
- UV radiation is harmful to eyes. Suitable eye protection must be worn when observing light output.
- Since this lamp operates at high voltages, extreme caution is require to prevent electrical shock.